How to adjust calibration parameter?

Uploaded:2012-00-14

1. In main operation window of the instrument, click the right key of the mouse, pops up the menu. Move the cursor to “Calibration/Manual Calibration. Adjust the calibration factors to 100%. Click the right  key  of  the  mouse, and pops  up  the  menu, then  move the cursor  to “Service/Maintenance”, enter the password 6666, and click “F12key.

2. Click   the   right  key   of  the   mouse,   pops   up  the   menu,  then   move   the   cursor  to

 

“Calibration/Factory Manual Calibration”. Adjust the calibration factors to 100%.

 

3. Use the calibration of Prokan recommended. In main operation window, measure the calibration. According to the measured histograms to adjust the gain of the passage. Adjust the gain of passage WBC, RBC according to the Multi-Parameter Assayed Hematology Control Expected Results Sheet.

4. In main operation window of the instrument, click the right key of the mouse, pops up the menu.


 then move the cursor to “Service/Test”, enter the password 6000,click “HGB Background/T”

 

key to adjust the potentiometer VR6,setting the voltage as 4.3±0.1V.